Testomat 2000®
Product number: 100011
4.662,70 €*
* Prices excl. VAT plus shipping costs.
Product information
Dimensions: 380 x 280 x 480 mm
Weight: 10,600 kg
Measurement parameters: Water hardness | Carbonate hardness | p-value | minus m-value
Measuring range: 0.89 - 448 ppm | 8.9 - 358 ppm | 0.1 - 15 mmol/l | 0.05 - 0.5 mmol/l
Range of application: Universal for water treatment plants
Menu navigation: German, English, French, Polish, Dutch, Italian, Spanish
Description
Description "Testomat 2000®"
The process photometer Testomat 2000® automatically determines water hardness, carbonate hardness, the p-value or the minus m-value via titration. It is ideal for monitoring the water quality of water treatment, water blending and drinking-water plants. Monitoring and quality control of softening plants with and without blending (Testomat 2000® reagent TH 2005–2250)Monitoring and quality control of decarbonisation plants (Testomat 2000® reagent TC 2050–2100) Measuring acidic values in decarbonisation and desalination plants (Testomat 2000® reagent TM 2005) Monitoring of steam boilers (Testomat 2000® reagent TP 2100).
Performance profile:
- Menu-driven operation and programming via plain text display
- Determinal measuring ranges of residual hardness, total hardness, carbonate hardness, minus m value, p value by indicator selection
- Selection of hardness units in ° dH, ° f, ppm CaCO3 or mmol / l
- High-precision titration using a piston dosing pump
- Analysis initiation:
- - Automatic interval operation (Interval pause adjustable from 0–99 min)
- - External control
- - Dynamic (fatigue-dependent interval operation)
- - Quantity dependent via contact water meter
- Two independent limits with hysteresis (1, 2 oder 3 bad analyzes) and adjustable switching functions
- Monitoring of two measuring points (Switching by external solenoid valves)
- Internal error documentation
- Programmable service address
- Indicator quantities: 100 and 500 ml bottles
- Programmable maintenance interval for maintenance request